Chapter 4
Column Oven
Oven Parameters
90
5. Scroll to
Final Temp 1
and enter the final temperature for the ramp.
6. Scroll to
Final Time 1
and enter the time the oven will maintain the Final
temp.
7. To end the single ramp program,
Ramp 2
must be
Off
.
OPERATING SEQUENCE
Setting Up Multiple Ramp Temperature Program
This program raises the initial oven temperature to a specified final temperature
through up to three ramps, each having a specified ramp rate, time, and
temperature.
Use
INC
and
DEC
keys to increase or decrease the parameter values. and confirm
pressing
ENTER
.
Figure 4-3. Example of Multiple Ramp Temperature Program
1. In Main Menu move the cursor until
Oven
is selected. Press
ENTER
to display
the
OVEN
menu.
2. Scroll to
Temp
and enter the initial the initial temperature.
100
200
300
°
C
10
20
30
40
min.
0
0
Initial
time
(9 min)
Final
time 2
(10 min)
Ramp 1
(10
°
C/min)
Final
temp 1
(150
°
C)
Final
temp 2
(300
°
C)
Ramp 2
(15
°
C/min)
Temp
(40
°
C)
Summary of Contents for FOCUS GC
Page 4: ...iv InstructionManual ...
Page 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Page 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Page 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Page 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Page 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Page 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Page 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Page 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Page 350: ...Index 350 Instruction Manual ...