Chapter 10
Autosamplers
Sequence Programming
196
When no vial
Skip - Abort
When Skip is set, the autosampler skips an eventual
missing sample vial and moves to find the next vial. The
sample sequence and the sample table of the data system
will be not affected. §
When Abort is set, the sequence will be aborted after
three missing vial
Start sequence
This is visualized when sequence is not running.
To start the sequence, select Start sequence, then press
ENTER
.
Stop sequence
This is visualized when sequence is running. To stop the
sequence, select Stop sequence, then press
ENTER
.
Table 10-4. Sequence Menu (Continued)
Parameter
Range
Description