Chapter 9
Thermal Conductivity Detector (TCD)
TCD Operating Modes
176
temperature limit in the TCD detector menu (Table 9-4). If this value is reached,
the system will automatically switch to the constant temperature mode.
The CV mode extends the linearity range to higher values, but with a negative
impact on sensitivity.
NOTE
Automatic Switching to Constant Current Mode
Every time the set values of filaments voltage, block temperature and filaments
temperature cause the filament current to reach the maximum value of 125 mA, the system
will automatically switch to the constant current mode and the filaments cannot be heated
more than the correspondent temperature.
This mode has good sensitivity and a linearity comparable to that obtained with the CV
mode. However, the high filament temperatures can potentially shorten the filament life.
Summary of Contents for FOCUS GC
Page 4: ...iv InstructionManual ...
Page 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Page 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Page 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Page 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Page 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Page 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Page 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Page 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Page 350: ...Index 350 Instruction Manual ...