Chapter 9
TCD Menu
Thermal Conductivity Detector (TCD)
179
TCD Menu
The
DETECTOR
menu contains the detector control parameters.
In Main Menu, move the cursor until
Detector
is selected. Press
ENTER
to
display the
DETECTOR
menu.
Constant Temperature Constant Voltage
DETECTOR (TCD)
DETECTOR (TCD)
Filament power Off
Fil status (not rdy)
Block temp 200 200
Trans temp 200 200
Cost fill temp?
Yes
Fil temp (--) 350
Reference On
Make-up On
Filament power Off
Fil status (not rdy)
Block temp 200 200
Trans temp 200 200
Cost fill temp? No
Fil volts (--) 10
Fil temp limit 350
Reference On
Make-up On
Table 9-4. The Detector (TCD) Parameters
Menu
Range
Comments
DETECTOR (TCD)
This line is the menu title bar.
Filament power
1
On/Off
Set
ON
to turn on the filament power. Set
OFF
to turn
off the filament.
Fil status
Ready/
Not Ready
This indicates the filament Ready or Not Ready
status.
Block temp
50–370 ºC
Set
ON
to turn on the detector heater and display the
actual and setpoint values.
Set
OFF
to turn off the heater.
Summary of Contents for FOCUS GC
Page 4: ...iv InstructionManual ...
Page 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Page 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Page 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Page 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Page 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Page 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Page 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Page 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Page 350: ...Index 350 Instruction Manual ...