Chapter 7
Column
Column Conditioning
156
Column Conditioning
To ensure optimum analytical performance, a new column must be conditioned to
remove contaminants and unwanted volatile compounds.
Column conditioning is required to remove volatile contaminants adsorbed from
the air.
Column conditioning consists of passing a carrier gas flow through the column
and heating the column to a temperature 20–50 °C above the maximum
temperature that will be used for running analyses, provided that temperature is
within the operating range of the column.
For detailed information on column conditioning of your specific column, refer to
the column manufacturer’s instructions.
CAUTION
With the column already installed, before starting column conditioning, disconnect
the column outlet from the detector. With the column not already installed, connect
only the column inlet to the injector.
WARNING! Do not use hydrogen as carrier gas because it could vent into the oven and present
an explosion hazard. Make sure that the system is leak tight before heating the
column oven. The column might be irreversibly damaged by the presence of
molecular oxygen at high temperature.
Summary of Contents for FOCUS GC
Page 4: ...iv InstructionManual ...
Page 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Page 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Page 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Page 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Page 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Page 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Page 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Page 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Page 350: ...Index 350 Instruction Manual ...