Chapter 7
Capillary and Wide-bore Columns
Column
127
Retaining Nuts
M4 split retaining nuts are used to connect capillary columns to injector and
detector base bodies. The nuts are split to allow easy installation and removal.
Figure 7-1 shows how to connect capillary or wide bore column to injector and
detector base body.
Figure 7-1. Capillary/Wide Bore Column to Injector and Detector Base Body Connections
Table 7-1. Ferrules
Capillary Column
Graphite Ferrules
0.1 mm ID
0.25 mm ID
0.25 mm ID
0.35 mm ID
0.32 mm ID
0.45 mm ID
0.53 mm ID
0.8 mm ID
1. Capillary/Wide bore Column
2. Graphite Ferrule
3. M4 Split Retaining Nut
1
2
3
INJECTOR or DETECTOR