88
8-8-16. Off timer value setting (
:
TIMER
:
SET)
This command sets or inquires about the Off Timer value.
Setting
:
TIMER
:
SET <value>
The <value> range is between 0.1 and 99.5. Specify the hour
in the integer part and minutes in the first decimal place (in
units of 10 minutes).
Application
example
:
TIMER
:
SET 10.3
Set the Off Timer value to 10 hours and 30 minutes.
Query
:
TIMER
:
SET?
Inquires about the setting.
Response
example
15.2
Indicates that the Off Timer value is set to 15 hours and 20
minutes.
Remark
This command causes an error during sequence operation
(RUN).
It also causes an error if a value between 6 and 9 is set in the
first decimal place.
8-8-17. Delay function ON/OFF (
:
DELAY
:
MOD)
This command activates or deactivates the delay function.
Setting
:
DELAY
:
MOD <value>
<value> is as shown below:
0: Deactivates the delay function. (Normal)
1: Activates the delay function.
Application
example
:
DELAY
:
MOD 1
Activates the delay function.
Query
:
DELAY
:
MOD?
Inquires about the setting.
Response
example
1
Indicates that the delay function is working.
“?” has the same
function as <value>.
Remark
This command causes an error during sequence operation
(RUN).