84
8-8-8. Hi-
Ω
time setting (
:
CONF
:
HIZ
:
HOLD)
This command sets the Hi-
Ω
function holding time in minutes.
Setting
:
CONF
:
HIZ
:
HOLD <value>
<value> is set in 5 steps between 5 and 30. Any value not
ending with 0 or 5 is rounded off.
Application
example
:
CONF
:
HIZ
:
HOLD20
Sets the Hi-
Ω
holding time to 20 minutes.
Query
:
CONF
:
HIZ
:
HOLD?
Inquires about the setting.
Response
example
5
Indicates that the Hi-
Ω
holding time is five minutes.
“?” has
the same function as <value>.
Remark
None
8-8-9. Display switching (
:
CONF
:
DISP)
This command switches the voltage, current and power display modes.
Setting
:
CONF
:
DISP <value>
<value> is as shown below:
1: CHA v current display
2: CHA v power display
3: CHA power + current display
4: CHB v current display
5: CHB v power display
6: CHB power + current display
7: Sequence or Off Timer display
For a one-output model, use the values 1 to 3 or 7 to specify a
display mode.
Application
example
:
CONF
:
DISP 1
Selects the v current display mode.
Query
:
CONF
:
DISP?
Inquires about the display setting.
Response
example
1
Indicates that the v current display mode is selected.
“?”
has the same function as <value>.
Remark
A value between 4 and 6 causes an error if it is set for a
one-output model.
The value of 7 causes an error unless the Off Timer or
sequence function (Pause, RUN) is working.