111
Functions
Constant-voltage (CV)
control with external voltage
Output voltage: Approx. 0V to 80V
for external voltage: 0V to 10V
Constant-voltage (CC)
control with external
resistance
Output voltage: Approx. 0V to 80V
for external resistance: 0
Ω
to 10K
Ω
Constant-current (CC) control
with external voltage
Output current: Approx. 0A to 120A (1200L) or
0A to 160A (1600L)
for external voltage: 0V to 10V
Constant-current (CC) control
with external resistance
Output current: Approx. 0A to 120A (1200L) or
0A to 160A (1600L)
for external resistance: 0
Ω
to 10k
Ω
Protective functions
Over-voltage protection
(OVP): Fixed
※
15
Output Off when output voltage exceeds 110% of
rated voltage.
Over-voltage protection
(OVP): Variable
※
16
Presettable in range from 1V to 84V on front panel.
Output Off when OVP works.
Over-current protection
(OCP): Fixed
※
15
Output Off when output current exceeds 110% of
rated current.
Over-current protection
(OCP): Variable
※
16
Presettable in range from 3A - 126A(1200L) / 2A
-168A(1600L) on front panel.
Output Off when OCP works.
Overheat protection (OHP)
※
15
Output Off at the internal heat sink temperature over
the set temperature.
NOTES
※
15
:
Throw the POWER switch again to reset.
※
16
:
Throw the POWER switch again or press the ESC key to reset.
Applicable Standards
The PSF-1200L / 1600L does not apply CE/LVD/EMC standards.
Others
Outside dimensions
(Projections not included.)
420mm(W)
×
124mm(H)
×
290mm(D)
Weight
PSF-1200L: Approx. 12kg