8-9-5. Clear and reset statuses
It is possible to cause clearing or resetting in the PSF Series power
supply unit by issuing commands or executing specific operations.
■
IFC (Interface clear)
Responses to the universal command IFC are as shown below:
・
The specified talker or listener status is cleared.
・
The GP-IB buffer, output queue and standing-by command remain
unchanged.
・
The SRQ remains unchanged.
・
The remote status and LLO setting remain unchanged.
・
Panel setting remains unchanged.
■
DCL SDC (Device clear)
Responses to the universal command DCL and address command
SDC are as shown below:
・
The GP-IB buffer, output queue and standing-by command are
cleared.
・
The interface status (specified talker or listener status) is cleared.
・
The SRQ and status byte are cleared.
・
The remote status and LLO setting are cleared and the power
supply unit enters the local state.
・
Panel setting remains unchanged.
■
*
RST (Reset command)
Responses to the IEEE488.2 command *RST are as shown below:
・
The specified talker or listener status is cleared.
・
The GP-IB buffer, output queue and standing-by command are
cleared.
・
The SRQ, status byte and mask setting are cleared.
・
The remote status and LLO setting remain unchanged.
・
Panel setting remains unchanged.
■
Turning on power again
When power is turned on again, the statuses are as shown below:
・
The specified talker or listener status is cleared.
・
The GP-IB buffer, output queue and standing-by command are
cleared.
・
The SRQ, status byte and mask setting are cleared.
・
The unit enters the local status.
LLO set before turning off power is cleared.
・
Panel setting at the time when power is turned off remains
unchanged.