79
8-8-1. Output voltage setting (
:
VOLT)
This command sets or inquires about the output voltage.
Setting
:
VOLT<value>
Sets the output voltage of a one-output model.
:
VOLT
:
A<value>
Sets the output voltage of CHA.
:
VOLT
:
B<value>
Sets the output voltage of CHB.
The <value> range is from 0.00 to 82.00.
Two decimal places are valid.
Application
example
:
VOLT 10.10
Sets the output voltage of a one-output model to 10.10V.
Query
:
VOLT?
Inquires about the set output voltage of a one-output model.
:
VOLT
:
A?
Inquires about the set output voltage of CHA.
:
VOLT
:
B?
Inquires about the set output voltage of CHB.
Response
example
10.10
Indicates that the set output voltage is 10.10V.
Remark
In tracking operation, setting of the CHA reflects upon the CHB
and CHB setting causes an execution error.
8-8-2. OVP setting (
:
VOLT
:
PROT)
This command sets or inquires about the OVP (over-voltage protection)
value.
Setting
:
VOLT
:
PROT<value>
Sets the OVP value of one-output model.
:
VOLT
:
PROT
:
A<value>
Sets the OVP value of CHA.
:
VOLT
:
PROT
:
B<value>
Sets the OVP value of CHB.
The <value> range is from 1.00 to 84.00.
Two decimal places are valid.
Application
example
:
VOLT: PROT 10.10
Sets the OVP value of a one-output model to 10.10V.
Query
:
VOLT
:
PROT?
Inquires about the set OVP value of a one-output model.
:
VOLT
:
PROT
:
A?
Inquires about the set OVP value of CHA.
:
VOLT
:
PROT
:
B?
Inquires about the set OVP value of CHB.
Response
example
10.10
Indicates that the set OVP value is 10.10V.
Remark
None