PRISM MPI2-25 and MPX2-25 Media Analysis Platform User Manual
56
Table 8:
Possible status in the VIDEO IP PORT, SFP Module line
Type of SFP installed
Data Rate setting
10Gb setting
25Gb setting
SFP 25Gb module
SFP not compatible…
(blank)
SFP 10Gb module
(blank)
SFP not compatible…
Unreadable MSA
data
Unknown
Unknown
No SFP installed
Disconnected
Disconnected
5.
Select
RS-FEC
if Forward Error Correction (FEC) is needed. RS-FEC is
specified by IEEE 802.3 Clause 91. (See
Figure 42
.)
NOTE.
FEC is only available when the Data Rate is set to 25Gb. FEC can only
be selected for both ports. If it is on for one, it is on for both.
Figure 42: VIDEO IP PORT submenu
1.
o
The IP Status display has the Data Rate setting, the FEC setting,
and the signal status. (See
Figure 43
.)
o
The Data Rate setting (10 or 25Gb) is on the right of the network
traffic bar.
o
The FEC setting is below the data rate.
o
The signal status is on the left of the network traffic bar.
Figure 43: IP Status display port status