AFA4000/2
Airflow Monitor / p.158
13.3
Auto sash alarm indications
In an alarm condition, originating from within the fume hood, the current alarm displays
in the status window:
Alarm message
Description
Obstruction Detected
This is displayed when the sash beam is broken and when
the sash starts to close.
Remove the obstruction and press the
Mute
button to reset
the alarm.
Sash Fault
This is displayed when the sash does not close or open when
expected.
Check the sash drive for faults and press
Mute
to reset the
alarm.
Remove Sash Lock /
Cancel to Reset
This is displayed when the sash is at or above the calibrated
lock position for longer than the pre-set alarm time.
Press
Mute
or lower the sash to reset the alarm.