AFA4000/2
Airflow Monitor / p.68
When the motor drive has been installed, open / close the sash to ensure that the sash
can travel freely.
6.4
Sash light curtain
The Sash Light Curtain consists of three main components and associated wiring:
•
Sash Light Curtain Control Box
•
LED Transmitter
•
LED Receiver
The Sash Light Curtain Control Box is connected to the main Auto Sash Control Unit.
6.4.1
Installing the sash light curtain control box
1.
Fix the Sash Light Curtain Control Box onto the top of the fume hood, using the tabs
attached to the control box and the screws provided. Ensure that there is access to
the terminals to connect the wiring.
Figure 38: Sash Light Curtain Control Box
2.
Use the pre-assembled cables and connectors supplied to connect the Light Curtain
Control Box to the Auto Sash Control Unit. See the wiring schematic in section 6.10
and Figure 31.
Summary of Contents for AFA4000/2
Page 1: ...AFA4000 2 Airflow monitor Installation and operating manual Issue 1 ...
Page 50: ...AFA4000 2 Airflow Monitor p 50 Figure 21 AFA4000 2 typical connection diagram ...
Page 51: ...AFA4000 2 Airflow Monitor p 51 Figure 22 AFA4000 2 AS typical connection diagram ...
Page 52: ...AFA4000 2 Airflow Monitor p 52 5 4 Dimensions Figure 23 AFA4000 2 and AFA4000 2 AS dimensions ...
Page 53: ...AFA4000 2 Airflow Monitor p 53 Figure 24 SM7 airflow sensor dimensions ...
Page 90: ...AFA4000 2 Airflow Monitor p 90 6 10 Auto Sash System Control Box general wiring diagram ...
Page 127: ...AFA4000 2 Airflow Monitor p 127 ...
Page 140: ...AFA4000 2 Airflow Monitor p 140 Figure 69 Typical connection diagram ...
Page 144: ...AFA4000 2 Airflow Monitor p 144 Figure 70 Connection diagram with RS232 485 converter ...