AFA4000/2
Airflow Monitor / p.108
Menu item
Option
Instructions
Auto Dim Screen
Auto dim screen
Set the time to auto dim the screen.
Low Air Timer
Low air repeat time
Enable or disable the timer.
Low air repeat time
If the timer is enabled, set the time.
Low Air Relay
Low air relay
Select the required relay:
•
None
•
Output 1
–
Output 3
Allow inhibit
Enable or disable the inhibit screen
feature.
High Air Relay
High air relay
Select the required relay:
•
None
•
Output 1
–
Output 3
Alarm Disable
Relay
Alarm disable relay
Select the required relay:
•
None
•
Output 1
–
Output 3
Sash High
Repeat timer
Enable or disable the timer.
Sash high time
If the timer is enabled, set the time.
Sash high relay
Select the required relay:
•
None
•
Output 1
–
Output 3
High Low Relay
High/low relay
Select the required relay:
•
None
•
Output 1
–
Output 3
Night Set Back
Night set-back
Select from:
•
Maintain Low Air
•
Reduce Low Air
Summary of Contents for AFA4000/2
Page 1: ...AFA4000 2 Airflow monitor Installation and operating manual Issue 1 ...
Page 50: ...AFA4000 2 Airflow Monitor p 50 Figure 21 AFA4000 2 typical connection diagram ...
Page 51: ...AFA4000 2 Airflow Monitor p 51 Figure 22 AFA4000 2 AS typical connection diagram ...
Page 52: ...AFA4000 2 Airflow Monitor p 52 5 4 Dimensions Figure 23 AFA4000 2 and AFA4000 2 AS dimensions ...
Page 53: ...AFA4000 2 Airflow Monitor p 53 Figure 24 SM7 airflow sensor dimensions ...
Page 90: ...AFA4000 2 Airflow Monitor p 90 6 10 Auto Sash System Control Box general wiring diagram ...
Page 127: ...AFA4000 2 Airflow Monitor p 127 ...
Page 140: ...AFA4000 2 Airflow Monitor p 140 Figure 69 Typical connection diagram ...
Page 144: ...AFA4000 2 Airflow Monitor p 144 Figure 70 Connection diagram with RS232 485 converter ...