AFA4000/2
Airflow Monitor / p.110
Menu item
Option
Instructions
BACNet settings
See section 8.1.1.4.
Sensor Error
Options
Buzzer on / off
Enable or disable the buzzer.
Sensor error relay
Select the required relay:
•
None
•
Output 1
–
Output 3
Sash Warning
Timer
Sash warning time
Set the time.
Sash warn rpt alarm
Enable or disable the alarm.
Sash Warning
Timer (cont.)
Sash warn rpt time
If the alarm is enabled, set the repeat
time.
Factory Settings
Enter password
Use
↑
/
↓
to enter password, then press
Mute.
Restore factory setting
Press Mute to restore the factory
settings. Press
↓
to return to the
Monitor config menu
(section 8.1.1).
Done
8.1.1.1
Cal config menu
Note: Use
↑
/
↓
to increase or decrease parameter values, then press
Mute
to save the
new value.
Menu item
Instructions
Display Units
Select the required display units:
•
m/sec
•
fpm
Low Air Alarm
Set the low air flow alarm threshold.
Low Air Cutoff
Enable or disable the low air cut-off.
If enabled, set the low air flow threshold.
Summary of Contents for AFA4000/2
Page 1: ...AFA4000 2 Airflow monitor Installation and operating manual Issue 1 ...
Page 50: ...AFA4000 2 Airflow Monitor p 50 Figure 21 AFA4000 2 typical connection diagram ...
Page 51: ...AFA4000 2 Airflow Monitor p 51 Figure 22 AFA4000 2 AS typical connection diagram ...
Page 52: ...AFA4000 2 Airflow Monitor p 52 5 4 Dimensions Figure 23 AFA4000 2 and AFA4000 2 AS dimensions ...
Page 53: ...AFA4000 2 Airflow Monitor p 53 Figure 24 SM7 airflow sensor dimensions ...
Page 90: ...AFA4000 2 Airflow Monitor p 90 6 10 Auto Sash System Control Box general wiring diagram ...
Page 127: ...AFA4000 2 Airflow Monitor p 127 ...
Page 140: ...AFA4000 2 Airflow Monitor p 140 Figure 69 Typical connection diagram ...
Page 144: ...AFA4000 2 Airflow Monitor p 144 Figure 70 Connection diagram with RS232 485 converter ...