AFA4000/2
Airflow Monitor / p.16
3.3
Controls and indicators
3.3.1
LED Halo indicator
The alarm unit has a LED Halo indicator that changes colour:
•
Red
Alarm
•
Amber
Caution
•
Green
Safe
The Red Alarm LED Halo is illuminated when pushbutton I is set to
Fan On/Off
and is
switched off.
3.3.2
Audible alarm sounder
The AFA4000/2 has an audible sounder with local or remote mute facilities. You can
permanently disable the alarm in the
Cal Config
Menu.
The audible alarm is muted when
Fan Off
is selected on pushbutton I or when the
Setback or Alarm Disable functions are activated.
When the audible alarm is muted/ disabled, the mute icon displays at the bottom left
side of the screen.
When an input is set to the mute function the keypad
Mute
button is disabled and you
can only mute the alarm by using the selected mute input.
3.3.3
Pushbuttons
The AFA4000/2 has 3 menu configurable pushbuttons. Each pushbutton can be
configured to a different function. The pushbutton icon and status is shown on the
display above the pushbutton.
Pushbutton I:
Fan On/Off
Displayed when the pushbutton is set to
FAN
operation.
Ext Fan On/Off
Displayed when the pushbutton is set to
Extract Fan
operation.
Summary of Contents for AFA4000/2
Page 1: ...AFA4000 2 Airflow monitor Installation and operating manual Issue 1 ...
Page 50: ...AFA4000 2 Airflow Monitor p 50 Figure 21 AFA4000 2 typical connection diagram ...
Page 51: ...AFA4000 2 Airflow Monitor p 51 Figure 22 AFA4000 2 AS typical connection diagram ...
Page 52: ...AFA4000 2 Airflow Monitor p 52 5 4 Dimensions Figure 23 AFA4000 2 and AFA4000 2 AS dimensions ...
Page 53: ...AFA4000 2 Airflow Monitor p 53 Figure 24 SM7 airflow sensor dimensions ...
Page 90: ...AFA4000 2 Airflow Monitor p 90 6 10 Auto Sash System Control Box general wiring diagram ...
Page 127: ...AFA4000 2 Airflow Monitor p 127 ...
Page 140: ...AFA4000 2 Airflow Monitor p 140 Figure 69 Typical connection diagram ...
Page 144: ...AFA4000 2 Airflow Monitor p 144 Figure 70 Connection diagram with RS232 485 converter ...