AFA4000/2
Airflow Monitor / p.101
6.
Press
↓
button, select
11. Auto Lock
and press
Enter
.
7.
Set to
Enabled
and press
Enter
.
8.
Press
↓
button, select
11b. Lock Alarm
and press
Enter
.
9.
Press
↑/↓
buttons to set the required alarm delay time and press
Enter
.
10.
Press
↓
button, select
11c. Lock Pos Calibration
and press
Enter
.
Instructions will be displayed, for example: Lock the sash then move the sash to the
lowest lock position. Enter to accept. Cancel to abort
11.
Open the sash to the Lock position and press
Enter
.
If the position chosen is too low, lower than the calibrated Bottom, Middle or Top
positions,
Bad position - try again
will be displayed.
If the chosen position is acceptable, the lock position will be set.
12.
Press
↓
button, select
Back to Set up menu
and press
Enter
.
13.
Press
↓
button, select
Exit and Save Changes
and press
Enter
.
The Auto Sash Controller will now go to
RUN
mode.
7.4.4
Calibrating the safety light curtain
To adjust the range and sensitivity of the Safety Light Curtain:
1.
Ensure the area between the Safety Light Curtain transmitter and receiver is free from
obstructions.
2.
Ensure the light beam paths are not broken and the Personnel Sensor is disabled so
that the sash does not Auto-Close whilst the Safety Light Curtain is being calibrated.
Note: If it is not possible to disable the Personnel Sensor, remove the motor / clutch
connector from the Auto Sash Control Unit. This activates the Sash Control
Alarm after the selected time delay and the reset button must be pressed once
the Safety Light Curtain has been calibrated.
3.
Remove the cover from the Safety Light Curtain Control Box.
Summary of Contents for AFA4000/2
Page 1: ...AFA4000 2 Airflow monitor Installation and operating manual Issue 1 ...
Page 50: ...AFA4000 2 Airflow Monitor p 50 Figure 21 AFA4000 2 typical connection diagram ...
Page 51: ...AFA4000 2 Airflow Monitor p 51 Figure 22 AFA4000 2 AS typical connection diagram ...
Page 52: ...AFA4000 2 Airflow Monitor p 52 5 4 Dimensions Figure 23 AFA4000 2 and AFA4000 2 AS dimensions ...
Page 53: ...AFA4000 2 Airflow Monitor p 53 Figure 24 SM7 airflow sensor dimensions ...
Page 90: ...AFA4000 2 Airflow Monitor p 90 6 10 Auto Sash System Control Box general wiring diagram ...
Page 127: ...AFA4000 2 Airflow Monitor p 127 ...
Page 140: ...AFA4000 2 Airflow Monitor p 140 Figure 69 Typical connection diagram ...
Page 144: ...AFA4000 2 Airflow Monitor p 144 Figure 70 Connection diagram with RS232 485 converter ...