AFA4000/2
Airflow Monitor / p.88
6.9.4
Personnel sensor dip switch settings
The following Personnel Sensor functions and parameters are controlled by Dip switches,
located within the Sash Light Curtain Control Box (Figure 62) (See section 7.4.7).
Figure 62: Personnel Sensor Dip Switch Settings
Switches
Setting
Description
1 & 2
Presence timer
Select the time delay before the Personnel Sensor
‘re
-
learns’ the background whe
n a stationary
object is detected.
2, 15, 60 or 180 second delays are selected by
setting associated switches 1 and 2 to
combinations of ON/OFF (Figure 62).
Note: The timer will reset if any movement is
detected during the time period.
3 & 4
Pattern depth
Set the distance of the detection field away from
the sensor.