AFA4000/2
Airflow Monitor / p.129
8.3.2
Navigating the system menus
8.3.2.1
Keypad tones
Menu item
Option
Instructions
Keypad tones
Enable or disable the keypad
sounding when a key is pressed.
8.3.2.2
Sash 1 settings menu
Menu item
Option
Instructions
Sash 1 settings
Enabled
Choose the setting to permanently
enable/disable the Auto Sash
Controller motor drive:
•
No to disable the drive.
•
Yes to enable the drive.
Close Delay
Set the time delay before the sash
closes.
(0
–
3600 seconds)
Close Alarm
Set the time before the alarm activates
before the sash closes.
(0
–
10
seconds)
Auto Open
Enable or disable Auto Open:
•
Enabled
•
Disabled
•
Auto Close Only
Open Delay
Set the time delay before the sash
opens.
(0
–
3600 seconds)
Open Alarm
Set the time before the alarm activates
before the sash opens.
(0
–
10 seconds)
Summary of Contents for AFA4000/2
Page 1: ...AFA4000 2 Airflow monitor Installation and operating manual Issue 1 ...
Page 50: ...AFA4000 2 Airflow Monitor p 50 Figure 21 AFA4000 2 typical connection diagram ...
Page 51: ...AFA4000 2 Airflow Monitor p 51 Figure 22 AFA4000 2 AS typical connection diagram ...
Page 52: ...AFA4000 2 Airflow Monitor p 52 5 4 Dimensions Figure 23 AFA4000 2 and AFA4000 2 AS dimensions ...
Page 53: ...AFA4000 2 Airflow Monitor p 53 Figure 24 SM7 airflow sensor dimensions ...
Page 90: ...AFA4000 2 Airflow Monitor p 90 6 10 Auto Sash System Control Box general wiring diagram ...
Page 127: ...AFA4000 2 Airflow Monitor p 127 ...
Page 140: ...AFA4000 2 Airflow Monitor p 140 Figure 69 Typical connection diagram ...
Page 144: ...AFA4000 2 Airflow Monitor p 144 Figure 70 Connection diagram with RS232 485 converter ...