AFA4000/2
Airflow Monitor / p.156
13.2
Diagnostics menu
The AFA4000 Diagnostics Menu can be used to check the comms settings and operation,
when the AFA4000 comms parameter settings have been configured.
Press the
↑
/
↓
buttons together from the Run screen to access the Diagnostics Menu. The
menu contains the following Input and Output data:
Menu
Parameter/Action
Alarm Test
The audible alarm will sound.
Comms Data
Displays the following:
•
Protocol (None / TEL / Modbus / BACnet)
•
Slave ID (for Modbus) or Device Instance (for BACnet)
•
Baud Rate
•
Parity
•
Tx
–
Number of Data Packets transmitted (increments when
transmitting). The displayed value rolls over to zero when
the maximum count is reached.
•
Rx
–
Number of Data Packets received (increments when
receiving). The displayed value rolls over to zero when the
maximum count is reached.
•
Tx & Rx = 0. AFA4000 is not connected to the network
(initial set up).
•
Tx & Rx >0, but fixed values. AFA4000 has lost connection
to the network.
Rx is counting but Tx is a fixed value. AFA4000 is connected to
the network but the Master (BMS) is offline or not polling the
AFA4000.
I/O Status
Input Data. This displays the following:
•
Input 1
–
On / Off or Analogue Voltage status
•
Input 2
–
On / Off or Analogue Voltage status
•
Input 3
–
On / Off or Analogue voltage status
Output Data. This displays the following:
Summary of Contents for AFA4000/2
Page 1: ...AFA4000 2 Airflow monitor Installation and operating manual Issue 1 ...
Page 50: ...AFA4000 2 Airflow Monitor p 50 Figure 21 AFA4000 2 typical connection diagram ...
Page 51: ...AFA4000 2 Airflow Monitor p 51 Figure 22 AFA4000 2 AS typical connection diagram ...
Page 52: ...AFA4000 2 Airflow Monitor p 52 5 4 Dimensions Figure 23 AFA4000 2 and AFA4000 2 AS dimensions ...
Page 53: ...AFA4000 2 Airflow Monitor p 53 Figure 24 SM7 airflow sensor dimensions ...
Page 90: ...AFA4000 2 Airflow Monitor p 90 6 10 Auto Sash System Control Box general wiring diagram ...
Page 127: ...AFA4000 2 Airflow Monitor p 127 ...
Page 140: ...AFA4000 2 Airflow Monitor p 140 Figure 69 Typical connection diagram ...
Page 144: ...AFA4000 2 Airflow Monitor p 144 Figure 70 Connection diagram with RS232 485 converter ...