AFA4000/2
Airflow Monitor / p.97
1.
From the
Run Screen
press and hold the
Mute
button for 5 seconds until the
Main
Menu
is displayed.
Note: When you go from the main menu to the run screen, a note briefly appears
saying ‘calibration required’.
2.
Use the
↑
/
↓
buttons to select
Set Up Auto Sash
. The display shows the message
TEL
SASH CLOSER
.
3.
Press the
Mute
button to display the current Hardware and Software version, for
example,
1. HW2.0 FW1.1
.
4.
Press the
↓
button to scroll down and select
Setup Menu
and press
Mute
.
5.
Enter the password (the factory default password is 1-0-0-1) and press
Mute
.
6.
In the
Setup
menu the controller will display the message
System Profile
.
7.
Use the
↓
button to scroll through the menu and select
Sash 1 Settings
, then press
Mute
.
8.
In the
S1 Settings menu
the display will show
1. Enabled
.
9.
Use the
↓
button to scroll through the menu and select
11. Auto Lock
. Press
Mute
.
10.
Set to
Enabled
and press
Mute
.
11.
Use the
↓
button to scroll through the menu and select
11b. Lock Alarm
. Press
Mute
.
12.
Use the
↑
/
↓
buttons to set the required alarm delay time. Press
Mute
.
13.
Use the
↓
button to scroll through the menu and select
11c. Lock Pos Calibration
.
Press
Mute
.
14.
The display will show on-screen instructions.
15.
Open the sash to the Lock position and press
Mute
.
When the selected position is too low, for example, it is lower than the calibrated
Lower, Middle or Top positions, the message
Bad position - try again
displays.
If the chosen position is OK, the lock position will be set.
16.
Use the
↓
button and select
Back to Set up menu
. Press
Mute
.
17.
Use the
↓
button and select
Exit and Save Changes
. Press
Mute
.
The controller returns to the Main Menu. Either select
Run
, or wait a few seconds for it to
automatically enter Run mode.
Summary of Contents for AFA4000/2
Page 1: ...AFA4000 2 Airflow monitor Installation and operating manual Issue 1 ...
Page 50: ...AFA4000 2 Airflow Monitor p 50 Figure 21 AFA4000 2 typical connection diagram ...
Page 51: ...AFA4000 2 Airflow Monitor p 51 Figure 22 AFA4000 2 AS typical connection diagram ...
Page 52: ...AFA4000 2 Airflow Monitor p 52 5 4 Dimensions Figure 23 AFA4000 2 and AFA4000 2 AS dimensions ...
Page 53: ...AFA4000 2 Airflow Monitor p 53 Figure 24 SM7 airflow sensor dimensions ...
Page 90: ...AFA4000 2 Airflow Monitor p 90 6 10 Auto Sash System Control Box general wiring diagram ...
Page 127: ...AFA4000 2 Airflow Monitor p 127 ...
Page 140: ...AFA4000 2 Airflow Monitor p 140 Figure 69 Typical connection diagram ...
Page 144: ...AFA4000 2 Airflow Monitor p 144 Figure 70 Connection diagram with RS232 485 converter ...