Figure 27: Auto Sash Control Unit Connectors, Single and Dual Fume Hoods
Once the Auto Sash Control Unit has been positioned, mark its position using the fixing
holes in the rear of the housing as a template. Then, attach the Auto Sash Control Unit
using screws.
6.1.2
Wiring in the Auto Sash Control Unit
For fume hoods with an inner and outer sidewall, route all cables through the sidewall
and into the service void.
For fume hoods with a single wall, route all cables up the outside of the sidewall, using
suitable trunking or fixing pads and ties.
When wiring in the Auto Sash Control Unit, observe the following:
•
Keep cables clear of moving parts.
•
Use grommets wherever cables pass through sidewalls to prevent abrasion of the
cable insulation and / or sidewall. Where a cable is supplied with a DIN connector, the
grommet should have a minimum internal diameter of 17mm (0.67").
•
Do not overstress cables by bending, for example when passing at 90° through a
sidewall.
Summary of Contents for AFA4000/2
Page 1: ...AFA4000 2 Airflow monitor Installation and operating manual Issue 1 ...
Page 50: ...AFA4000 2 Airflow Monitor p 50 Figure 21 AFA4000 2 typical connection diagram ...
Page 51: ...AFA4000 2 Airflow Monitor p 51 Figure 22 AFA4000 2 AS typical connection diagram ...
Page 52: ...AFA4000 2 Airflow Monitor p 52 5 4 Dimensions Figure 23 AFA4000 2 and AFA4000 2 AS dimensions ...
Page 53: ...AFA4000 2 Airflow Monitor p 53 Figure 24 SM7 airflow sensor dimensions ...
Page 90: ...AFA4000 2 Airflow Monitor p 90 6 10 Auto Sash System Control Box general wiring diagram ...
Page 127: ...AFA4000 2 Airflow Monitor p 127 ...
Page 140: ...AFA4000 2 Airflow Monitor p 140 Figure 69 Typical connection diagram ...
Page 144: ...AFA4000 2 Airflow Monitor p 144 Figure 70 Connection diagram with RS232 485 converter ...