AFA4000/2
Airflow Monitor / p.155
Sash Fault Alarm
always shown
The controller does not detect that the sash position sensor
or limit switch is fitted, check the connections.
Fault LED is illuminated
Ensure that the area between the sash sensor transmitter
and receiver is clear and that the sensors are aligned, power
up again to reset.