AFA4000/2
Airflow Monitor / p.146
12.
Operation: Auto Sash Controller
12.1
Control philosophy
When all components are installed, setup and calibrated correctly, the Auto Sash
Controller closes the sash after a pre-set delay, provided that the area in front of the
fume hood is clear.
The following additional protocols also apply:
•
When the user returns to the fume hood during the delay period, the control timer
will stop and reset once the user leaves the fume hood again.
•
When the user returns to a fume hood that is closing, the drive will stop when it
senses the user.
•
When an obstruction is detected under the sash, the Auto Sash Controller will not
open or close the sash and will give a visual alarm to indicate a fault condition.
•
When an obstruction is detected whilst the sash is closing, the sash drive will stop and
give a visual indication to demonstrate a fault condition. To reset the fault alarm,
press the reset button.
•
When the sash starts to close a timer is initiated. When the sash does not close within
the specified time, the sash drive will stop and will give a visual alarm to indicate a
fault condition.
•
When a user is present in front of the fume hood, the sash can be operated by hand.
12.2
Modes of operation
The Auto Sash Controller has the following modes of operation:
•
User Present
•
User Not Detected
•
Auto-Open (User Configurable)
•
Building Management System (BMS) Inputs
•
Sash Lock
Summary of Contents for AFA4000/2
Page 1: ...AFA4000 2 Airflow monitor Installation and operating manual Issue 1 ...
Page 50: ...AFA4000 2 Airflow Monitor p 50 Figure 21 AFA4000 2 typical connection diagram ...
Page 51: ...AFA4000 2 Airflow Monitor p 51 Figure 22 AFA4000 2 AS typical connection diagram ...
Page 52: ...AFA4000 2 Airflow Monitor p 52 5 4 Dimensions Figure 23 AFA4000 2 and AFA4000 2 AS dimensions ...
Page 53: ...AFA4000 2 Airflow Monitor p 53 Figure 24 SM7 airflow sensor dimensions ...
Page 90: ...AFA4000 2 Airflow Monitor p 90 6 10 Auto Sash System Control Box general wiring diagram ...
Page 127: ...AFA4000 2 Airflow Monitor p 127 ...
Page 140: ...AFA4000 2 Airflow Monitor p 140 Figure 69 Typical connection diagram ...
Page 144: ...AFA4000 2 Airflow Monitor p 144 Figure 70 Connection diagram with RS232 485 converter ...