AFA4000/2
Airflow Monitor / p.81
Figure 54: Sash position sensor
2.
Mark the fixing holes on the framework and attach the sensor.
The fixing bracket has two 0.37 mm (0.015") diameter fixing holes.
Note: Ensure that the sensor is in a position that does not over extend the draw wire.
The sensor draw wire full stroke is 1270 mm (50") maximum.
3.
Attach the draw wire to either:
•
The sash using a suitably sized screw, the draw wire has a 9.53 mm (0.38") diameter
fixing ring with a 4.85 mm (0.18") diameter fixing hole in its centre.
•
The sash cable using a cable tie, ensuring that the fixing ring does not run over the
pulleys.
•
The counterweight using a cable (Figure 55).