AFA4000/2
Airflow Monitor / p.168
15.2.3
Controller connections
Connector
Terminal Connection
Ratings &
Specification
Power Supply CN15
•
2 m power lead with flying
lead with 3 Way plug in
terminal block
•
0.75mm2, 300 V, 6 A Rated
Cable
•
100-240 Vac; 50/60
Hz; 3A (Fused 3.15
A)
Keypad CN4
•
Backlit LCD display with
pushbuttons
•
RJ45 Cable 2 m
•
Max 24 Vdc 100
mA (PTC Fused 200
mA)
Drive Motor Assembly CN7
•
2 m flying lead with 4-way
plug-in terminal block
•
Motor 0-24 Vdc
PWM Output
Clutch 24 Vdc
•
Max total power
135W (5.6A)
Personnel Sensor CN6
•
2 m flying lead with 4-way
plug-in terminal block
•
Max 24 Vdc 100
mA (PTC Fused
200 mA)
Sash Position Sensor CN5
•
2 m flying lead with 3-way
plug-in terminal block
•
Bottom Switch: 24
V (PTC Fused
200 mA)
•
Sash Pos Sensor:
3.3 Vdc 43 mA Max
Sash Sensor CN3
•
PIR sensor with Glass
Detection Light Curtain or
Under Sash Sensor with 3
m flying lead
•
Max 24 Vdc 100
mA (PTC Fused
200 mA)
Tilt Switch CN2
•
2-way plug in terminal
block
•
Max 24 Vdc 100
mA (PTC Fused
200 mA)
Summary of Contents for AFA4000/2
Page 1: ...AFA4000 2 Airflow monitor Installation and operating manual Issue 1 ...
Page 50: ...AFA4000 2 Airflow Monitor p 50 Figure 21 AFA4000 2 typical connection diagram ...
Page 51: ...AFA4000 2 Airflow Monitor p 51 Figure 22 AFA4000 2 AS typical connection diagram ...
Page 52: ...AFA4000 2 Airflow Monitor p 52 5 4 Dimensions Figure 23 AFA4000 2 and AFA4000 2 AS dimensions ...
Page 53: ...AFA4000 2 Airflow Monitor p 53 Figure 24 SM7 airflow sensor dimensions ...
Page 90: ...AFA4000 2 Airflow Monitor p 90 6 10 Auto Sash System Control Box general wiring diagram ...
Page 127: ...AFA4000 2 Airflow Monitor p 127 ...
Page 140: ...AFA4000 2 Airflow Monitor p 140 Figure 69 Typical connection diagram ...
Page 144: ...AFA4000 2 Airflow Monitor p 144 Figure 70 Connection diagram with RS232 485 converter ...