AFA4000/2
Airflow Monitor / p.150
Current status
Description
Sash Inhibited
Displayed when the tilt switch input is open. The drive is
inhibited until the tilt switch is closed.
Sash Disabled
Displayed when the sash drive has been disabled in the
menu settings.
Sash Locked
Displayed when the sash is at or above the calibrated lock
position.
12.3.2
Auto Sash settings
12.3.2.1
Stand alone
Auto-Open can be set to
Disabled, Enabled
or
Enabled if Auto-Closed.
When the Auto-Open feature is enabled in the menu and is switched on from the
keypad, an
*
displays in the left-hand corner of the screen. For example,
* User present
.
To enable Auto-Open, press and hold the
Mute
button for 5 seconds or until the
*
is
displayed.
To disable Auto-Open, press and hold the
Mute
button for 5 seconds or until the
*
disappears.
Note: If set to Enabled if Auto-Closed, the sash will only Auto-Open if the sash has Auto-
Closed.
Closing the sash manually or by using the pushbuttons, will inhibit the Auto-Open when
the user returns to the fume hood.
12.3.2.2
AFA4000/2/AS
The Auto Sash can be set to be
Disabled
,
Enabled
or
USR enabled
(
User enable / disable
).
To User enable the Auto Sash, press and hold the
Mute
and
↑
buttons for 1 second.
To User disable the Auto Sash, press and hold the
Mute
and
↓
buttons for 1 second.
USR
disabled
is displayed in the Status window.
Note: Auto open can be set to disabled / enabled / enabled if auto closed. When set to
enabled and when auto closed, the sash will only Auto open when the sash has
Summary of Contents for AFA4000/2
Page 1: ...AFA4000 2 Airflow monitor Installation and operating manual Issue 1 ...
Page 50: ...AFA4000 2 Airflow Monitor p 50 Figure 21 AFA4000 2 typical connection diagram ...
Page 51: ...AFA4000 2 Airflow Monitor p 51 Figure 22 AFA4000 2 AS typical connection diagram ...
Page 52: ...AFA4000 2 Airflow Monitor p 52 5 4 Dimensions Figure 23 AFA4000 2 and AFA4000 2 AS dimensions ...
Page 53: ...AFA4000 2 Airflow Monitor p 53 Figure 24 SM7 airflow sensor dimensions ...
Page 90: ...AFA4000 2 Airflow Monitor p 90 6 10 Auto Sash System Control Box general wiring diagram ...
Page 127: ...AFA4000 2 Airflow Monitor p 127 ...
Page 140: ...AFA4000 2 Airflow Monitor p 140 Figure 69 Typical connection diagram ...
Page 144: ...AFA4000 2 Airflow Monitor p 144 Figure 70 Connection diagram with RS232 485 converter ...