Objects are detected when infra-red light from the transmitter is not detected by the
receiver due to an obstruction in its path.
Figure 9: Sash Light Curtain Transmitter and Receiver
A separate Sash Light Curtain Control Box includes a sensitivity potentiometer that
allows the beam sensitivity adjusted to sense glass or other objects that are partially
transparent to infra-red light.
Figure 10: Sash Light Curtain Control Box
4.3.5
Under Sash sensor
The Under Sash Sensor is an alternative to the Sash Light Curtain which detects whether
the opening below the sash is clear from obstructions before the sash is closed.
A sensor on one side of the fume hood detects the reflection of a beam of light from a
self-adhesive strip of retro-reflective tape mounted on the opposite inside wall of the
fume hood.