AFA4000/2
Airflow Monitor / p.37
4.3.2
Personnel sensor
Figure 5: Detection area clear
Figure 6: Obstruction present
The Personnel Sensor is a Passive Infra-Red (PIR) occupancy detector which detects the
presence / absence of the fume hood user.
Note: If the user is detected, the sash will not automatically close. The Sash can be
opened / closed manually or using the
Up
/
Down
pushbuttons.
On power up, the Personnel Sensor ‘learns’ the reflection characteristics of the
environment within its field of view and stores this information as reference background
data. The sensor then 're-learns' the background every three minutes. A stationary
object, such as a stool, that is left in the field becomes part of the background and is
'ignored' by the sensor. When the user is not present at the fume hood, the sash will
close automatically. This follows a pre-set delay time, if the Sash Light Curtain is not
broken and the sash is open.
Green LED
Red LED
Summary of Contents for AFA4000/2
Page 1: ...AFA4000 2 Airflow monitor Installation and operating manual Issue 1 ...
Page 50: ...AFA4000 2 Airflow Monitor p 50 Figure 21 AFA4000 2 typical connection diagram ...
Page 51: ...AFA4000 2 Airflow Monitor p 51 Figure 22 AFA4000 2 AS typical connection diagram ...
Page 52: ...AFA4000 2 Airflow Monitor p 52 5 4 Dimensions Figure 23 AFA4000 2 and AFA4000 2 AS dimensions ...
Page 53: ...AFA4000 2 Airflow Monitor p 53 Figure 24 SM7 airflow sensor dimensions ...
Page 90: ...AFA4000 2 Airflow Monitor p 90 6 10 Auto Sash System Control Box general wiring diagram ...
Page 127: ...AFA4000 2 Airflow Monitor p 127 ...
Page 140: ...AFA4000 2 Airflow Monitor p 140 Figure 69 Typical connection diagram ...
Page 144: ...AFA4000 2 Airflow Monitor p 144 Figure 70 Connection diagram with RS232 485 converter ...