AFA4000/2
Airflow Monitor / p.20
3.5
External connections
3.5.1
Inputs
The AFA4000/2 has the following inputs:
3.5.1.1
Input 1, Input2, Input 3
These are volt-free inputs configurable for:
•
Normally closed relays
•
Normally open relays or
•
Analogue 0 - 5 Vdc input
The inputs can be configured as:
Digital input functions (closed or
open volt free contact):
Alarm disabled
Night set-back
External alarm
Emergency
Sash high
High / Low
Sash warning
Mains fail
Fan stop
Mute
Fire Alarm
Analogue input functions:
Temperature
3.5.2
Outputs
The AFA4000/2 has the following outputs:
3.5.2.1
Relay outputs 1, 2, 3
These are volt-free outputs configurable for:
•
Normally closed relays or
•
Normally open relays
3.5.2.2
Analogue output 1
0 - 10 Vdc re-transmission of face velocity over 0 - 1.00 m/sec, for example 0.50 m/sec =
5 Vdc.
Summary of Contents for AFA4000/2
Page 1: ...AFA4000 2 Airflow monitor Installation and operating manual Issue 1 ...
Page 50: ...AFA4000 2 Airflow Monitor p 50 Figure 21 AFA4000 2 typical connection diagram ...
Page 51: ...AFA4000 2 Airflow Monitor p 51 Figure 22 AFA4000 2 AS typical connection diagram ...
Page 52: ...AFA4000 2 Airflow Monitor p 52 5 4 Dimensions Figure 23 AFA4000 2 and AFA4000 2 AS dimensions ...
Page 53: ...AFA4000 2 Airflow Monitor p 53 Figure 24 SM7 airflow sensor dimensions ...
Page 90: ...AFA4000 2 Airflow Monitor p 90 6 10 Auto Sash System Control Box general wiring diagram ...
Page 127: ...AFA4000 2 Airflow Monitor p 127 ...
Page 140: ...AFA4000 2 Airflow Monitor p 140 Figure 69 Typical connection diagram ...
Page 144: ...AFA4000 2 Airflow Monitor p 144 Figure 70 Connection diagram with RS232 485 converter ...