AFA4000/2
Airflow Monitor / p.133
8.3.2.4
BMS output settings
Menu item
Option
Instructions
BMS Output
Settings
Alarm Sounder
Enable / disable the remote alarm
sounder.
Analog Volts
Select the required analogue output
voltage for sash position indication:
•
0
–
10V
•
1
–
10V
•
0
–
5V
•
1
–
5V
Analog Current
Select the required analogue output
current for sash position indication:
•
0
–
20mA
•
4
–
20mA
Relays 1
–
4
(RLY1, RLY2, RLY3,
RLY4)
Select the required settings for each
of the four relays:
•
Always Open
•
Always Closed
•
Open on Sash Alarm
•
Close on Sash Alarm
•
Open on Obstruction Detected
•
Close on Obstruction Detected
•
Open on User Detected
•
Close on User Detected
•
Open on Sash Open
•
Close on Sash Open
•
Open on Sash Locked
•
Close on Sash Locked
Summary of Contents for AFA4000/2
Page 1: ...AFA4000 2 Airflow monitor Installation and operating manual Issue 1 ...
Page 50: ...AFA4000 2 Airflow Monitor p 50 Figure 21 AFA4000 2 typical connection diagram ...
Page 51: ...AFA4000 2 Airflow Monitor p 51 Figure 22 AFA4000 2 AS typical connection diagram ...
Page 52: ...AFA4000 2 Airflow Monitor p 52 5 4 Dimensions Figure 23 AFA4000 2 and AFA4000 2 AS dimensions ...
Page 53: ...AFA4000 2 Airflow Monitor p 53 Figure 24 SM7 airflow sensor dimensions ...
Page 90: ...AFA4000 2 Airflow Monitor p 90 6 10 Auto Sash System Control Box general wiring diagram ...
Page 127: ...AFA4000 2 Airflow Monitor p 127 ...
Page 140: ...AFA4000 2 Airflow Monitor p 140 Figure 69 Typical connection diagram ...
Page 144: ...AFA4000 2 Airflow Monitor p 144 Figure 70 Connection diagram with RS232 485 converter ...