Performing measurements
R&S
®
NRPM
79
Mannual 1425.8663.02 ─ 08
Trigger Level
Sets the trigger threshold for internal triggering derived from the test signal.
Remote command:
Trigger Delay
Sets the delay between the trigger event and the actual start of the measurement.
Remote command:
Dropout
With a positive (negative) trigger slope, the dropout time is the minimum time for
which the signal must be below (above) the power level defined by
.
Remote command:
Holdoff
Sets the hold-off time.
After a trigger event, the sensor module suppresses all trigger events that occur
within this time period.
Remote command:
Hysteresis
Sets the hysteresis in dB. A trigger event occurs, if the trigger level:
●
Falls below the set value on a rising slope.
●
Rises above the set value on a falling slope.
Thus, you can use this setting to eliminate the effects of noise in the signal for the
edge detector of the trigger system.
Remote command:
7.2.2.6
System settings
Provides the parameters of the general network environment and specific identifi-
cation parameters of the sensor module in the network.
Measurement applications