Table 9-13: Meaning of bits used in the operation sense status register
Bit No.
Meaning
0
Not used
1
sensor module initializing
The sensor module is being initialized.
5 to 15
Not used
9.2.8.5
Operation lower limit fail status register
The
CONDition
register contains information about whether a displayed value is
currently below a configured lower limit. The
EVENt
register indicates whether a
measured value dropped below a limit value since the last readout of the Opera-
tion Lower Limit Fail Status Register. Details of the behavior are defined by the
transition register.
Querying the register:
●
STATus:OPERation:LLFail:CONDition?
●
STATus:OPERation:LLFail[:SUMMary][:EVENt]?
Table 9-14: Meaning of bits used in the operation lower limit fail status register
Bit No.
Meaning
0
Not used
1
Lower limit fail
The measured value drops below the lower limit value.
5 to 15
Not used
9.2.8.6
Operation upper limit fail status register
The
CONDition
register contains information about whether a displayed value is
currently above a configured upper limit. The
EVENt
register indicates whether a
limit value was exceeded since the last readout of the Operation Upper Limit Fail
Status Register.
Querying the register:
●
STATus:OPERation:ULFail:CONDition?
●
STATus:OPERation:ULFail[:SUMMary][:EVENt]?