Parameters:
<state>
*RST:
ON
Manual operation:
10.8.2
Setting the frequency
The frequency of the signal to be measured is not automatically determined. For
achieving better accuracy, the carrier frequency of the applied signal must be set.
..............................................................................168
[SENSe<Sensor>:]FREQuency
<frequency>
Transfers the carrier frequency of the RF signal to be measured. This frequency
is used for the frequency-response correction of the measurement result.
The center frequency is set for broadband signals, e.g. spread-spectrum signals,
multicarrier signals.
The data of the connected antenna module determines the frequency limits.
Parameters:
<frequency>
Range:
antenna-specific
*RST:
10.0e9
Default unit: Hz
Example:
FREQuency 4.2e10
Manual operation:
10.8.3
Configuring corrections
It is possible to set some parameters that compensate for a change of the mea-
sured signal due to fixed external influences.
................................................................................... 169
.......................................................................................... 169
Configuring basic measurement parameters