Firmware update
R&S
®
NRPM
88
Mannual 1425.8663.02 ─ 08
The size of the file is 10242884. This number has 8 digits. Thus, the
<block_data>
consist of the following:
●
#
●
8
How many digits follow to specify the file size.
●
10242884
Number that specifies the file size.
●
<file_contents>
Contents of the
*.rsu
file, byte-by-byte
●
0x0a
Delimiter
In this example, you write exactly 10242905
bytes to the sensor module, for
example by using a 'viWrite()' function.
The 10242905
bytes result from the values of the list above:
9 + 1 + 1 + 1 + 8 + 10 1
In a (pseudo) string notation, the memory block looks as follows:
SYST:FWUP #810242884<file_contents>0x0a