Remote control commands
R&S
®
NRPM
148
Mannual 1425.8663.02 ─ 08
Example: Average count = 1
[SENSe<Sensor>:]TRACe:AVERage:COUNt
For average count 1, the setting of the termination control has no impact. In both
cases, the measurement runs for the duration of one trace time. Then, settled
trace data are available, and the sensor module returns to the idle state.
1
2
3
4
5
1 = Trigger event
2 = Noninverted chopper phase
3 = Trace measurement result
4 = Trigger synchronization
5 = Return to idle state
10.6
Selecting a measurement mode and retrieving
results
Before starting a measurement, select the measurement mode using:
.
The following modes are available:
●
Continuous average (
"POWer:AVG"
): After a trigger event, the power is inte-
grated over a time interval, see
Chapter 10.7.1, "Configuring a continuous
●
Trace (
"XTIMe:POWer"
): A sequence of measurements is performed, see
Chapter 10.7.3, "Configuring a trace measurement"
After measuring, you can query the measurement results with the correspondent
FETCh
command.
Selecting a measurement mode and retrieving results