Remote control commands
R&S
®
NRPM
157
Mannual 1425.8663.02 ─ 08
[SENSe<Sensor>:]LIST:FREQuency
...................................................................... 157
[SENSe<Sensor>:]LIST:FREQuency:POINts?
.........................................................157
CONTrol<Sensor>:FREQuency:MODE
.................................................................. 157
[SENSe<Sensor>:]LIST:FREQuency
<frequency>...
Sets the frequency values for list processing mode.
Parameters:
<frequency>
Range:
0 to 110e9
*RST:
10e9
Default unit: Frequency
Example:
[SENSe1:]LIST:FREQuency
40e09,45e09,58e09,75e09
[SENSe<Sensor>:]LIST:FREQuency:POINts?
Sets the number (points) of frequency entries in list processing mode.
Example:
[SENSe1:]LIST:FREQuency:POINts 4
Usage:
Query only
CONTrol<Sensor>:FREQuency:MODE
<mode>
Selects the frequency list mode.
Parameters:
<mode>
FIXed | RXList
*RST:
FIXed
Example:
CONTrol1:FREQuency:MODE RXList
10.7.3
Configuring a trace measurement
The trace measurement is used to acquire the course of power over a certain
time. During the measurement time (
) a large
number of measurements are made and the result is returned to the user as an
array of values with a predefined size
length of an individual measurement(-point) is determined from the ratio of mea-
Configuring the measurement modes