Remote control commands
R&S
®
NRPM
162
Mannual 1425.8663.02 ─ 08
Example:
C1Af3260xxxxyyyy
C1 Channel 1
A The letter ‘A’ to denote
the Average-Trace
f The letter ‘f’ to denote
float format
3 3 bytes for length of the number
of points that follows
260 260 float values (4 bytes each)
xxxxyyyy... 260 float values...
This example shows one of up to nine measurement data
blocks which could be contained in the user data content.
Referring to the example above, if the min and max traces
are also selected (by
SENS:AUX MINMAX
) the subsequent
data would contain:
C1mf3260mmmmnnnn
C1 Channel 1
m The letter ‘m’ to denote
the Min-Trace
f The letter ‘f’ to denote
float format
3 3 bytes for length of the
number of points that follows
260 260 float values (4 bytes each)
mmmmnnnn... 260 float values...
C1Mf3260gggghhhh
C1 Channel 1
M The letter ‘M’ to denote
the Max-Trace
f The letter ‘f’ to denote
float format
3 3 bytes for length of the
number of points that follows
260 260 float values (4 bytes each)
gggghhhh... 260 float values...
For further information, see
the
...[NRPM3]\Trace-M3\Visa\
visaTraceBinaryData-M3
project from the accompany-
ing programming examples (included). The source-code
shows how to extract the measurement data from the
"user-data-content".