Remote control commands
R&S
®
NRPM
147
Mannual 1425.8663.02 ─ 08
Example: Moving termination control
Further settings for this example:
●
[SENSe<Sensor>:]TRACe:AVERage:TCONtrol
Every chopper phase is started by a trigger event and lasts the defined trace
time. During a chopper phase, the trigger synchronization is high (
). Every measurement provides a result. After 2 measurements,
when the trace average count is reached, settled trace data result is available.
1
2
3
4
5
6
7
8
1 = Start of the measurement cycle
2 = Trigger event
3 = Noninverted chopper phase
4 = Inverted chopper phase
5 = Trigger synchronization
6 = Averaged trace data result after trace average count is reached
7 = Trace measurement result before trace average count is reached
8 = Return to the start of the measurement cycle
Controlling the measurement