Remote control commands
R&S
®
NRPM
180
Mannual 1425.8663.02 ─ 08
Manual operation:
TRIGger:SYNC:PORT
<sync_port>
Selects the external connection for the sensor module's sync output, see also
.
Parameters:
<sync_port>
EXT1 | EXTernal1 | EXT2 | EXTernal2
*RST:
EXT1
TRIGger:SYNC:STATe
<state>
Usually used in combination with
If activated, blocks the external trigger bus as long as the sensor modules
remains in the measurement state.
This function makes sure, that a new measurement only starts after all sensor
modules have completed their last measurement.
Make sure that the number of repetitions is the same for all the sensor modules
involved in the measurement. Otherwise, the trigger bus is blocked by any sensor
module that has completed its measurements before other sensor module and
has returned to the IDLE state.
Parameters:
<state>
*RST:
OFF
10.11
Using the status register
Further information:
●
Chapter 9.2, "Status reporting system"
Contents:
General status register commands
............................................................... 181
Reading out the CONDition part
................................................................... 181
..........................................................................182
Using the status register