Remote control commands
R&S
®
NRPM
134
Mannual 1425.8663.02 ─ 08
SYSTem[:SENSor]:NAME
<sensorname>
Queries the sensor name, or assigns an arbitrary alias name to the sensor.
If not specified, it defaults to the hostname of the sensor module.
Parameters:
<sensorname>
Example:
SYSTem[:SENSor]:NAME
"SensorModuleInput_A90D"
SYSTem[:SENSor]:NAME?
Queries the sensor name.
Response:
"SensorModuleInput_A90D"
Manual operation:
10.4.2
Handling of available antenna modules
The
SENSe:CHANnel
subsystem contains commands for checking connected
antenna modules, and activating individual antenna modules.
Remote commands:
[SENSe<Sensor>:]CHANnel<Channel>:PRESence?
...............................................134
[SENSe<Sensor>:]CHANnel<Channel>[:ENABle]
................................................... 134
[SENSe<Sensor>:]CHANnel<Channel>:PRESence?
Queries whether an antenna module is connected to a channel.
Suffix:
<Channel>
.
1...3
Usage:
Query only
Manual operation:
[SENSe<Sensor>:]CHANnel<Channel>[:ENABle]
<state>
Deactivates channels without an antenna module connected.
Note:
A sensor module channel which does not have a connected antenna mod-
ule causes a static error condition. The static error leads to a blinking LED. There-
fore we recommend that you disable unconnected channels by this command.
Configuring the general functions