Performing measurements
R&S
®
NRPM
75
Mannual 1425.8663.02 ─ 08
The remote commands required to define the sensor settings are described
in:
●
Chapter 10.4.2, "Handling of available antenna modules"
●
Chapter 10.8.3, "Configuring corrections"
●
Chapter 10.13, "Calibrating/zeroing the R&S
●
Chapter 10.12, "Testing the R&S
NRPM OTA Power Measurement Solu-
............................................................................................... 75
Antenna
Activates an antenna module module for the measurement.
Remote command:
[SENSe<Sensor>:]CHANnel<Channel>:PRESence?
[SENSe<Sensor>:]CHANnel<Channel>[:ENABle]
Zero Calibration
Executes zeroing using the signal at the sensor module input, see
Note:
Turn off all test signals before zeroing. An active test signal during zeroing causes
an error.
Remote command:
CALibration<Channel>:ZERO:AUTO
Sensor Information
Provides information on the R&S
NRPM OTA power measurement system.
Measurement applications