Remote control commands
R&S
®
NRPM
186
Mannual 1425.8663.02 ─ 08
CALibration<1...4>:DATA:LENGth?
Queries the length in bytes of the calibration data set currently stored in the flash
memory. Programs that read out the calibration data set can use this information
to determine the capacity of the buffer memory required.
Suffix:
<1...4>
.
1...4
Usage:
Query only
CALibration<Channel>:ZERO:AUTO
<state>
Performs zeroing using the signal at the sensor module input.
Disconnect the sensor module from all power sources, since any signal at the RF
input of the sensor module affects the calibration.
While zero calibration is in progress, no queries or other setting commands are
allowed. Any communication attempt can run into a timeout.
The setting command accepts only the parameter
ONCE
;
OFF
and
ON
are ignored.
The query returns the value
ON
if a calibration is in progress, otherwise the value
OFF
.
Suffix:
<Channel>
.
1...4
The channel suffix is ignored.
Parameters:
<state>
*RST:
OFF
Manual operation:
Calibrating/zeroing the R&S
NRPM3(N) sensor module