Remote control commands
R&S
®
NRPM
143
Mannual 1425.8663.02 ─ 08
Example: Repeating termination control
Further settings for this example:
●
[SENSe<Sensor>:]AVERage:TCONtrol
The measurement is started by the trigger event. Due to the chopper phases, one
measurement lasts twice the defined aperture time. As defined by the average
count, after 4 measurements, the result is averaged and available. During the
whole measurement cycle, the trigger synchronization is high (
1
2
3
4
5
6
7
8
1 = Start of the measurement cycle
2 = Trigger event
3 = Noninverted chopper phase
4 = Inverted chopper phase
5 = Duration of one aperture time (1 x t
AP
)
≙
length of one chopper phase
6 = Measurement result
7 = Trigger synchronization
8 = Return to the start of the measurement cycle
Controlling the measurement