Remote control commands
R&S
®
NRPM
169
Mannual 1425.8663.02 ─ 08
10.8.3.1
Duty cycle corrections
The duty cycle is the percentage of one period in which the signal is active, when
pulse-modulated signals are corrected. The duty cycle is only evaluated in the
Continuous Average mode.
Remote commands:
[SENSe<Sensor>:]CORRection:DCYCle
................................................................ 169
[SENSe<Sensor>:]CORRection:DCYCle:STATe
......................................................169
[SENSe<Sensor>:]CORRection:DCYCle
<duty_cycle>
Effective in continuous average mode.
Sets the duty cycle for measuring pulse-modulated signals. The duty cycle
defines the percentage of one period when the signal is active.
If the duty cycle is enabled, the R&S
NRPM calculates the signal pulse power
from the average power considering the duty cycle in percent.
Parameters:
<duty_cycle>
Range:
0.001 to 100.00
*RST:
1.00
Default unit: Percent
Example:
CORRection:DCYCle 0.01
Manual operation:
[SENSe<Sensor>:]CORRection:DCYCle:STATe
<state>
Activates the duty cycle correction for the measured value.
Parameters:
<state>
*RST:
OFF
Example:
CORRection:DCYCle:STATe ON
Manual operation:
10.8.3.2
Offset corrections
The offset accounts for external losses by adding a fixed level offset in dB.