background image

1

© 2019 Eddyfi Technologies

TECHNICAL DOCUMENTATION |

M2M

GEKKO

32:128 | 64:64 | 64:128

Revision A.7

Redacted by Victor FOSSARD

Approved by Thierry COUTURIER

Date :  21/06/2019

Summary of Contents for GEKKO M2M

Page 1: ...1 2019 Eddyfi Technologies TECHNICAL DOCUMENTATION M2M GEKKO 32 128 64 64 64 128 Revision A 7 Redacted by Victor FOSSARD Approved by Thierry COUTURIER Date 21 06 2019 ...

Page 2: ...2 PHASED ARRAY CONNECTOR 3 UT CONNECTORS 4 ENCODER CONNECTOR 5 I O CONNECTOR 6 POWER CONNECTOR 7 BATTERY SLOT 8 AUTOMATIC BOOT 9 FRONT PANEL 10 ACCESSORIES 11 GLOBAL WARNING 12 BATTERY WARNING 13 CLEANING 14 TECHNICAL SPECIFICATIONS ...

Page 3: ...3 1 MECHANICAL DRAWING ...

Page 4: ...CH ADAPT Connector Utility Connector Information Supplier I PEX Reference 30046 160T F Matching Connector Supplier I PEX References straight 30056 160T F right angle 30047 160T 04F For electric safety reasons only accessories approved by Eddyfi Technologies can be used with our system Before purchasing any probe please contact us 2 PHASED ARRAY CONNECTOR ...

Page 5: ...sed array channel number 1 to 128 Phased array probe channel 1 to 128 GND Ground pin For better ultrasound result all GND pin have to be connected to probe ground Connector Signal Description 2 PHASED ARRAY CONNECTOR Depend on the hardware configuration 32 128PR 64 64PR or 64 128PR ...

Page 6: ... Connector Utility Connector Mapping Connector Information Supplier LEMO Reference ERN 00 250 CTL Matching Connector UT connector is NimCAMAC standard Supplier LEMO Reference FGG 00 302 CLAD35 3 UT CONNECTORS For electric safety reasons only accessories approved by EDDYFI can be used with our system Before purchasing any probe please contact us ...

Page 7: ...de Number of available encoders 2 or 3 depending on software setup and option INPUTS OUTPUTS 2 TTL inputs Connector Information Supplier Connector type LEMO EGG Reference EGG 1K 316 CLL Matching Cable Supplier LEMO Connector Male 16 Pin Reference FGG 1K 316 CLAC65Z 4 ENCODER CONNECTOR Connector Utility ...

Page 8: ...Mapping 4 ENCODER CONNECTOR 5V Encoder input Buffer 470Ω Encoder input Front View of the Connector Signal Name Description User matching signal Encoder Phase A B TTL 5V Absolute Max current 50 mA Max frequency 4MHz ...

Page 9: ... 5V 100ns pulse generated with each Pulser signal IPEX connector Signal can be used to synchronize an external system Jitter 10ns Absolute Max output current 50 mA If configured as INPUT see in Capture software as below Type 1 axis encoder increment TTL 5V active on High level Max Input current 50µA When connected to Phased Array Wheel probe Synchro1 is used as Index Increment short press and Scan...

Page 10: ...ement TTL 5V PRF 5V 100ns pulse generated with each Pulser signal IPEX connector Signal can be used to synchronize an external system Jitter 10ns Absolute Max output current 50 mA If configured as INPUT see in Capture software as below Type 1 axis encoder increment TTL 5V active on High level Max Input current 50µA When connected to Phased Array Wheel probe Synchro2 is used as Pause Resume to togg...

Page 11: ...tputs currently not driven by the software Development in progress Connector Information Supplier Connector type LEMO EGG Reference EGG 2K 326 CLL Matching Cable Supplier LEMO Connector Male 26 Pin Reference FGG 2K 326 CLAC65 Connector Utility ...

Page 12: ...l be automatically detected by the GEKKO system and will keep same screen resolution as GEKKO 1024x768 ETHERNET When plugged to a network the GEKKO appears with the name GEKKKO XXXX and the Capture files folder is shared with anybody Refer to EDDYFI procedure PRO_GEKKO_INS_022 SHARING USER GUIDE PRO_GEKKO_INS_027_GEKKO_V2_Connectivity_US A The ethernet controller is 1Gb s speed When the I O hatch ...

Page 13: ...ttery will be charged by the external power supply Only use the external power supply supplied by EDDYFI with the GEKKO When plugging the external power supply and if there are no batteries GEKKO automatically boots up Connector Information Supplier FISCHER Reference DEU 104 A037 130 Matching Cable Do not use another external power supply Supplier EDDYFI Reference CAB_0048 POWER GEKKO 6 POWER CONN...

Page 14: ...14 Connector Mapping 6 POWER CONNECTOR When the power cable is plugged the position of the system should allow to easily unlock the connector This is in order to shut down the unit in case of emergency ...

Page 15: ...r charges the lower battery at first When the two batteries have the same charge level they are charged at the same time It takes 6 hours to completely charge 2 flat batteries WITHOUT WALL ADAPTOR BATTERIES POWER THE GEKKO Gekko can work with one battery only If batteries have different charge levels the Gekko drains more power from the most charged battery until the levels are equal Then GEKKO dr...

Page 16: ...used to AUTOMATICALLY and REMOTELY boot the unit up To AVOID AUTOMATIC BOOT of the GEKKO while inserting the batteries without plugging any external power supply or while plugging the external power supply without any batteries inserted Press Battery Status Button Without external power supply WHEN external power is off or unplugged Inserting a battery in the system automatically boots the unit BA...

Page 17: ...the battery level even if the system is off ON OFF Push this button to power ON the system Press on ON OFF button briefly to shut down properly the GEKKO Only in case the system is frozen you can push the ON OFF button for 4 seconds to force the power down ON OFF To AVOID AUTOMATIC BOOT of the GEKKO while inserting the batteries without plugging any external power supply or while plugging the exte...

Page 18: ...ont Panel View 9 FRONT PANEL CLICKABLE SCROLL WHEEL TURN the wheel to adjust values cursors PRESS the wheel to validate the new value position OTHERS For other buttons see software user guide CLICKABLE SCROLL WHEEL ...

Page 19: ...n Picture Easy Latch EDDYFI ref IMP_0061 EASYLATCH ADAPT This accessory allows the connection of a probe with an EASY_LATCH connector to GEKKO system Dongle Wi Fi EDIMAX ref EW 7811Un This accessory allows the WiFi connection to the GEKKO system ...

Page 20: ...n of a probe with an HYPERTRONICS FRB connector to the GEKKO system Available splitter models SPLITTER 1x64 2x32 SPLITTER 1x128 4x32 SPLITTER 1x128 2x64 SPLITTER 1x64 2x30 4 LEMO SPLITTER 1x128 2x62 4 LEMO EDDYFI ref CAB_0109 SPL FRB128 4X32 The GEKKO channels are splitted to two different I PEX connectors and LEMO 00 connectors optional ...

Page 21: ...cture BATTERY EDDYFI ref ADD BATT NI2020 Additional battery 10 8V Lithium Ion 8 7Ah 94Wh BATTERY CHARGER EDDYFI ref CHARG 2BATT CH5050A External dual battery charger able to charge 2 GEKKO batteries in 4hours 10 ACCESSORIES_BATTERIES ...

Page 22: ...cription Picture Adaptor for MicroD25 system EDDYFI ref CAB_0069D This cable allows to connect MicroD25 encoder type to GEKKO system Adaptor for SUBD15 system EDDYFI ref CAB_0104A This cable allows to connect SUBD15 encoder type to GEKKO system ...

Page 23: ...tive conductor of the power cord sector The plug must be inserted only into a mains socket outlet with ground contact You should never cancel function protection using an extension cord power cable without a protective conductor grounding When the protective grounding seems damaged you must power down the unit and prevent unintentional operation The device must only be connected to a power source ...

Page 24: ...35 V Do not insert your finger inside two sink fins on the back of the system Do not insert your finger between back rubber part and axis needed for carrying strap The GEKKO system have V E S A compatible fixation on the back Please contact EDDYFI before using these fixation plot Do not insert your finger in the two batteries slot It is forbidden to handle the system by the top rubber part The sys...

Page 25: ...approved by EDDYFI NDT Do not store the battery if the battery level is below 30 Recharge the battery charge level between 30 and 70 before storing Remove the batteries from GEKKO when storing the device during a long time The transport of Li ion batteries is governed by the Recommendations on the transportation of dangerous goods in United States Governments intergovernmental organizations and ot...

Page 26: ...al appearance of the unit clean the housing with a soft cloth To remove stubborn stains use a damp cloth with a mild soap solution Do not use abrasives or strong solvents which could damage the finish Once you have removed the connector protection make sure they are dry before connecting anything else If the connectors are wet dry them with a soft dry cloth or let them air dry Never use abrasives ...

Page 27: ...64PR or 64 128PR E N V I R O N M E N T IP rating Designed for IP66 Operating temperature 10 to 45 C 14 to 113 F Storage temperature w batteries 20 to 60 C 4 to 140 F w o batteries 20 to 70 C 4 to 158 F Drop tested According to MIL STD 810G C O N N E C T I V I T Y Phased Array IPEX x1 UT TOFD LEMO 00 x4 x4 P R or x2 P C Encoder Input LEMO16 connector Up to 3 TTL Quadrature or clock dir 4MHz max Syn...

Page 28: ...Up to 64 active channels simultaneously Input impedance 50Ω Frequency range Frequency range 0 4 to 20 MHz Max input signal 2 Vpp Gain 0 to 120 dB 0 1dB step Active aperture Up to 64 elements Compliant with EN ISO 18563 1 F M C T F M TFM Real time up to 128 elements 256 Kpi Refresh rate Up to 110 Hz at 65 Kpi ATFM Adaptative TFM Real time FMC Recording capability Modes Direct indirect and converted...

Page 29: ...p Gain 0 to 120 dB 0 1dB step Compliant with EN ISO 12668 1 14 TECHNICAL SPECIFICATIONS Depends on the hardware configuration 32 128PR 64 64PR or 64 128PR D i g i t i z e r Amplitude resolution 16 bits Real time averaging 1 to 32 FMC A scan range 8K max Sampling frequency Up to 100MHz Signal processing Rectified RF Envelope A scan size Up to 65K samples A c q u i s i t i o n PRF Up to 40 KHz Data ...

Page 30: ...30 Contact us Any questions Support m2m eddyfi com Hardware problems with your unit MaintenanceEddyfiEurope eddyfi com ...

Page 31: ...31 ...