Remote control commands
R&S
®
NRPM
176
Mannual 1425.8663.02 ─ 08
TRIGger:HOLDoff
<holdoff>
Sets the hold off time, see
Parameters:
<holdoff>
Range:
0.00 to 10.00
*RST:
0.00
Default unit: s
Manual operation:
TRIGger:HYSTeresis
<hysteresis>
Sets the hysteresis thresholds. A trigger event occurs, if the trigger level:
●
Falls below the set value on a rising slope.
●
Rises above the set value on a falling slope.
You can use the hysteresis to eliminate the effects of noise in the signal on the
edge detector of the trigger system.
Parameters:
<hysteresis>
Range:
0.00 to 10.00
*RST:
0.00
Default unit: dB
Manual operation:
TRIGger:IMMediate
Triggers a generic trigger event that causes the sensor module to exit the
WAIT_FOR_TRIGGER state immediately, irrespective of the trigger source and
the trigger delay and start the measurement. The command is the only means of
starting a measurement when the trigger source is set to HOLD. Only one mea-
surement cycle is executed irrespective of the averaging factor.
Usage:
Event
TRIGger:LEVel
<level>
Sets the trigger threshold for internal triggering derived from the test signal.
This setting is effective only for the internal
Configuring the trigger