Remote control commands
R&S
®
NRPM
174
Mannual 1425.8663.02 ─ 08
TRIGger:ATRigger[:STATe]
<state>
Activates the artificial trigger.
An artificial trigger is generated if the time set with
has elapsed after the start of measurement and no trigger event has occurred.
Parameters:
<state>
*RST:
OFF
Example:
TRIG:ATR:STAT ON
TRIGger:COUNt
<count>
Sets the number of measurement cycles to be performed when the measurement
is started. This number equals the number of results which can be obtained from
the sensor module after a single
. As long as the
defined number of measurements are not yet executed, the sensor module auto-
matically initiates another measurement internally after the current result is availa-
ble.
Use this command in particular in conjunction with buffered measurements, for
example, to fill a buffer with a predefined size with measurements which have
been triggered with the command
or with the configured trigger condition.
Parameters:
<count>
Range:
1 to 8192
*RST:
1
TRIGger:DELay
<delay>
Sets the delay between the trigger event and the beginning of the actual mea-
surement (integration).
Parameters:
<delay>
Range:
-5.0 to 10.0
*RST:
0.0
Default unit: s
Manual operation:
Configuring the trigger