Remote control commands
R&S
®
NRPM
175
Mannual 1425.8663.02 ─ 08
TRIGger:DELay:AUTO
<state>
Activates the automatic setting of the delay time.
If activated, the measurement starts only after the R&S
NRPM input is settled.
The function determines the delay value automatically. It is ignored if the set
time is longer than the automatically determined value.
Parameters:
<state>
*RST:
OFF
Manual operation:
TRIGger:DTIMe
<dropout_time>
Sets the dropout time for the internal trigger source. During this time, the signal
power must exceed (negative trigger slope) or undercut (positive trigger slope)
the level defined by the trigger level
.
At least, this time must elapse before triggering can occur again.
See
Parameters:
<dropout_time>
Range:
0.00 to 10.00
*RST:
0.00
Default unit: s
Manual operation:
TRIGger:EXTernal<2...2>:IMPedance
<impedance>
Sets the termination resistance of the second external trigger input (
EXTernal2
).
You can select between
HIGH
(~ 10 kOhm) and
LOW
(50 Ohms) to fit the impe-
dance of the trigger source and thus minimize reflections on the trigger signals.
Suffix:
<2...2>
.
2...2
Parameters:
<impedance>
HIGH | LOW
*RST:
HIGH
Configuring the trigger